Description
Características
Escaner para Hi – Pot 5,000V – Hioki 3930
- » Automatic Multipoint Testing of High Voltages
- » Maximum of 32 Channels In single mode, an individual 3930 can measure eight channels, and up to four units can be connected together for a total of 32 channels.In multi-mode, the 3930 has four high and low channels, respectively;and when four units are connected, up to 16 high and low channels can be tested.
- » Safe Design The 3930 can handle up to 5kV of direct current or alternating current.
- » The 3930 is a general-purpose high voltage scanner that can not only be controlled by using the HIOKI 3153’s program function, but also by general-purpose logic and sequencers
- » Primary Applications Insulation withstand voltage testing for electrical and electronic equipment, Insulation withstand voltage testing for electric components
- » Operation modes Multi- and single modes ( External switch )
- » Number of channels Multi-mode; 4 high channels and 4 low channels, Single mode; 8 high channels and a common channel
- » Rated voltage used AC 5 kV/DC 5 kV
- » Control method: General-purpose control
- » Maximum open and closed voltage: 5000 V DC, 5000 V AC
- » Maximum open and closed current: 1.0 A (open and closed capacity: 50 W)
- » Contact point indirect contact resistance: 500 mor less, with 1 mA AC
- » Contact point maximum capacity: 50 W
- » Operation time: 6 ms or less
- » Recovery time: 6 ms or less
- » Degree of Accuracy: Standards for current leakage when applying voltage Single mode, no output cable, and all output relays Turned on for both AC and DC. When applying DC (1000 V); 0.1 µA or less/unit When applying AC (5 kV, 50/60 Hz only); 0.4 mAor less/unit (Differs depending on the status of the connection cable)
- » Operation temperature range: 0°C to 40°C, 80% rh or less (no condensation)
- » Storage temperature range: -10°C to 50°C, 90% rh or less (no condensation)
- » Operation environment: Indoors, altitude of 2000 m or less
- » Withstand voltage: High voltage terminal – between the chassis: AC 10 kV, 10 mA, 1 min
- » Power: VSCV 24 V DC, ±10% (applied using the control signal input connector)
- » Maximum rated power: 12 VA
- » Measurements: Approx. 316 (W) × 100 (H) × 350 (D) mm
Qué Incluye:
Connection cables H.V. TEST LEAD 9615
Reviews
There are no reviews yet.